MOC63u

Main Features


• Equipped with the UniBloc aluminum block, to provide accurate moisture measurements.

• Equipped with a high output halogen heater capable of rapid heating.

• Wide observation window to allow checks of sample status during heating.

• 60 g capacity/0.001 g minimum display

• The temperature on the pan can be set between 50 ?C and 200 ?C.

• Can store 10 sets of measurement conditions and 100 data items.

• Cleaning and maintenance are easy.

• Liquid and paste samples can also be measured using fiberglass sheets.

• Energy saving design (32 % reduction in comparison to previous Shimadzu models)


Sample Applications 


• Fiberglass sheets for liquid measurement were used to promote liquid 

evaporation.

• Two measurement conditions were used, timed ending and automatic 

ending modes. Essentially the same average values were obtained. 

With samples whose principal component has a relatively high 

evaporation temperature and also contains moisture, the same results 

will be obtained regardless of the mode used


Sample Measurement of Corn Starch 


• Approx. 5 g of corn starch was added to the pan, and was spread over 

the entire surface using the tip of a spoon.

• No change in appearance was evident after drying.

• Favorable repeatability of 1 % max. was obtained.